International Journal of Life science and Pharma Reviews (IJLPR)  
   
 
International Journal of Life science and Pharma Research (IJLPR)
Pharmaceutical Sciences
Volume 10 issue 2, April - June 2020    Pages:26-33
Quantification of Atazanavir and Ritonavir in Human Plasma Samples by Rp-Hplc Include Method of Detection in the Title, Eg: Using PDA Detection

ABHINANDANA.PATCHALA AND RAMARAO NADENDLA
[View PDF]
DOI: http://dx.doi.org/10.22376/ijpbs/lpr.2020.10.2.P 26-33
Abstract:
A fixed oral dose combination of Atazanavir and Ritonavir is currently used for the treatment of patients with HIV infections. A selective and novel bio-analytical technique was designed to evaluate Atazanavir and Ritonavir by mass spectroscopic investigation in plasma matrices. The method was chromatographed with Agilent TC-C18, 4.6 x 75 mm, 3.5 μm, 80 Å column, 5mM ammonium acetate: acetonitrile (20:80 v/v) mobile phase was used for Chromatographic separation. UV detector was used to detect the Atazanavir and Ritonavir at 235 nm. For extraction of the analyte and internal standard, Liquid-liquid extraction was employed .This method is validated over a linear concentration range of 50.0 – 10000.0 ng/ml for Atazanavir and Ritonavir with a correlation coefficient (r) of ≥ 0.9997 and both drugs were stable in plasma samples.
Keywords: HPLC; Atazanavir, Ritonavir, Human plasma; Bio analytical
 
 